Logo

ASTM F2503-08

Historical Revision

Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment

$83.00

$83.00

$149.40


Sub Total (1 Item(s))

$ 0.00

Estimated Shipping

$ 0.00

Total (Pre-Tax)

$ 0.00


...

p id="s00002">1.1 This practice covers the marking of medical devices and other items to indicate their safety in the magnetic resonance (MR) environment.

1.2 The purpose of this practice is to (1) recommend that items that may be brought into the MR environment be permanently marked to indicate the MR environment to which a specific item may safely be exposed, and (2) recommend information that should be included in the marking. It is recognized that direct marking on the item is not practical for implants and certain other medical devices. Where direct marking is not practical, this practice recommends that the marking be included in the labeling and on patient information cards (see 7.1).

1.3 Image artifact is not considered to be a safety issue and so is not addressed in this practice (see X1.5).

1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Medical devices and other items have caused serious injuries and death for patients and other individuals in the MR environment.

This practice provides a uniform system for marking to indicate the MR conditions that have been determined to be acceptable for a medical device or other item. It provides simple visual icons and terms which are intended to reduce injuries and other mishaps that occur when items that pose hazards in the MR environment are brought into the MR environment.

SDO ASTM: ASTM International
Document Number F2503
Publication Date Oct. 1, 2008
Language en - English
Page Count 7
Revision Level 08
Supercedes
Committee F04.15
Publish Date Document Id Type View
April 1, 2023 F2503-23E01 Revision
April 1, 2023 F2503-23 Revision
Feb. 1, 2020 F2503-20 Revision
June 1, 2013 F2503-13 Revision
Oct. 1, 2008 F2503-08 Revision
Aug. 1, 2005 F2503-05 Revision