1.1 This guide describes a laboratory method using a weight-loss technique for evaluating the wear properties of materials or devices, or both, which are being considered for use as bearing surfaces of human-hip-joint replacement prostheses. The hip prostheses are evaluated in a device intended to simulate the tribological conditions encountered in the human hip joint, for example, use of a fluid such as bovine serum, or equivalent pseudosynovial fluid shown to simulate wear mechanisms and debris generation as found in vivo, and test frequencies of 1 Hz or less.

1.2 Since the hip simulator method permits the use of actual implant designs, materials, and physiological load/motion combinations, it can represent a more physiological simulation than basic wear-screening tests, such as pinion-disk (see Practice F 732) or ring-on-disk (see ISO-6474).

1.3 It is the intent of this guide to rank the combination of implant designs and materials with regard to material wear-rates under simulated physiological conditions. It must be recognized, however, that there are many possible variations in the in vivo conditions, a single laboratory simulation with a fixed set of parameters may not be universally representative.

1.4 The reference materials for the comparative evaluation of candidate materials, new devices, or components, or a combination thereof, shall be the wear rate of extruded or Compression-molded, ultra-high molecular weight (UHMW) polyethylene (see Specification F 648) bearing against standard counter faces Stainless Steel (see Specification F 138); cobalt-chromium-molybdenum alloy (see Specification F 75); thermomechanically processed cobalt chrome (see Specification F 799); alumina ceramic (see Specification F 603), having typical prosthetic quality, surface finish, and geometry similar to those with established clinical history. These reference materials will be tested under the same wear conditions as the candidate materials.


SDO ASTM: ASTM International
Document Number F1714
Publication Date Sept. 10, 1996
Language en - English
Page Count 7
Revision Level 96
Supercedes
Committee F04.22
Publish Date Document Id Type View
Sept. 10, 1996 F1714-96R02 Revision
Sept. 10, 1996 F1714-96 Revision
April 1, 2018 F1714-96R18 Reaffirmation
March 15, 2013 F1714-96R13 Reaffirmation
June 1, 2008 F1714-96R08 Reaffirmation