1.1 This specification covers breathing systems and components employed with anesthesia gas machines for humans. This specification considers the circle system as a whole and some of its components individually. A particular emphasis is placed upon component arrangement in the circle absorber-type system, and submits a system of standard description and notation. Excluded are ventilators for use during anesthesia, Mapelson nonrebreathing type systems, as well as breathing systems and related components of dental analgesia machines. (For rationale, see Appendix X1.)

1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

1.3 This specification is arranged as follows:

SectionScopeReferenced DocumentsTerminologySystem ClassificationGraphic NotationsDiagramatic RulesFunctional Segments of the Circle SystemTest Procedures-GeneralSterilizationSystemsLeakage of Breathing SystemResistance of Breathing SystemsVolume of Gas Not Delivered to Patient Due to Internal ComplianceMonitoring Requirements
Adjustable Pressure-Limiting (APL) ValvesReservoir BagsCO2 AbsorbersUnidirectional ValvesRequirementsTest ProcedureOptional ComponentsBreathing TubesFresh Gas Inlet and Fresh Gas Supply Tube or HoseConnectorsPressure GaugesY-Piece and Breathing TubesInformation in LabelingRationale


This specification covers breathing systems and components employed with anesthesia gas machines for humans. This specification considers the circle system as a whole and some of its components individually. A particular emphasis is placed upon component arrangement in the circle absorber-type system, and submits a system of standard description and notation. The circle system is divided into four segments: segment 1 - from patient to expiratory valve, segment 2 - from expiratory valve to reservoir bag, segment 3 - from reservoir bag to inspiratory valve, and segment 4 - from inspiratory valve to patient. Leakage of breathing system and resistance of breathing system shall be tested to meet the requirements prescribed. The adjustable pressure-limiting valves, reservoir bag connectors and reservoir bags, carbon dioxide absorbers, unidirectional valves, and optional components shall meet the requirements prescribed.


SDO ASTM: ASTM International
Document Number F1208
Publication Date May 1, 2005
Language en - English
Page Count 13
Revision Level 89(2005)
Supercedes
Committee F29.10
Publish Date Document Id Type View
May 1, 2005 F1208-89R05 Revision
Jan. 1, 2000 F1208-89R00E01 Revision
Jan. 1, 2000 F1208-89R94E01 Revision
Jan. 1, 2000 F1208-89R00 Reaffirmation