1.1 This test method covers the measurement of soluble, suspended, and total 226Ra in water in concentrations above 3.7 × 10−3 Bq/L. This test method is not applicable to the measurement of other radium isotopes.

1.2 This test method may be used for quantitative measurements by calibrating with a 226Ra standard, or for relative measurements by comparing the measurements made with each other.

1.3 This test method does not meet the current requirements of Practice D2777.

1.4 The values stated in SI units are to be regarded as standard. The values given in parentheses are mathematical conversions to inch-pound units that are provided for information only and are not considered standard.

1.5 Hydrofluoric acid (HF) is very hazardous and should be used in a well-ventilated hood. Wear rubber gloves, safety glasses or goggles, and a laboratory coat. Avoid breathing any HF fumes. Clean up all spills promptly and wash thoroughly after using HF.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.


5.1 The most prevalent of the five radium isotopes in ground water, having a half life greater than one day, are 226Ra and 228Ra. These two isotopes also present the greatest health risk compared to the other naturally occurring nuclides of equal concentrations if ingested via the water pathway.

5.2 Although primarily utilized on a water medium, this technique may be applicable for the measurement of the 226Ra content of any medium once the medium has been completely decomposed and put into an aqueous solution.

5.3 This test method is based on a method previously published by Rushing, et al. (1).3 The general methodology and basis of the technique are similar to that of Ref (2).

SDO ASTM: ASTM International
Document Number D3454
Publication Date May 1, 2021
Language en - English
Page Count 9
Revision Level 21
Supercedes
Committee D19.04
Publish Date Document Id Type View
May 1, 2021 D3454-21 Revision
Oct. 1, 2018 D3454-18 Revision
June 15, 2011 D3454-11 Revision
Feb. 1, 2005 D3454-05 Revision
Aug. 10, 1997 D3454-97 Revision